Answer to Question #8095 Submitted to "Ask the Experts"

Category: Medical and Dental Equipment/Shielding — Shielding

The following question was answered by an expert in the appropriate field:

Q
I am looking for guidance on the requirements for labeling of lead-lined walls. Is it required by the Nuclear Regulatory Commission or any other body that each lead-lined wall in a radiology clinic be labeled with its lead equivalent? If so what what publication might I find it in?
A

Use of x-ray procedures in the United States is regulated by each individual state. Certain states may have a rule that requires the amount of shielding in each x-ray room wall be indicated in or near the room. As a consultant serving clients in four different states, I have only seen one facility that had postings indicating the amount of lead in each wall. Although it seems to be a good idea, it does not appear to be common practice. Review the regulations for the state you are interested in to see if there are any requirements.

Ken "Duke" Lovins, CHP
Health Physicist
 

Answer posted on 17 March 2009. The information posted on this web page is intended as general reference information only. Specific facts and circumstances may affect the applicability of concepts, materials, and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice. To the best of our knowledge, answers are correct at the time they are posted. Be advised that over time, requirements could change, new data could be made available, and Internet links could change, affecting the correctness of the answers. Answers are the professional opinions of the expert responding to each question; they do not necessarily represent the position of the Health Physics Society.